| Measurement objects |
Whole blood or hemolysis sample |
| Measured items |
Fast Mode: HbA1c (Stable HbA1c, S-A1c) and HbF
Variant Mode: HbA1c (Stable HbA1c, S-A1c) and HbF (HbS and HbC separation and HbE and HbD detection can be performed). |
| Measurement principle |
Reversed-phase cation exchange chromatography |
| Measurement wavelength |
420 nm/500 nm (Dual-wavelength colorimetry) |
| Resolution |
0.1% Ratio, 1 mmol/mol |
| Measurement range |
HbA1c: 3 – 20%, 9 – 195 mmol/mol
HbF: 0 – 100% |
| Processing speed |
Variant mode: 160 seconds/test
Fast mode: 100 seconds/test |
| Sample consumption |
Whole blood sample: Approximately 4µL
Anemia sample: Approximately 8 µL
Hemolysis sample: Approximately 350 µL |
| Sample container |
Sample tube:
(12.3 or 15 mm in outer diameter) × (75 to 100 mm in height)
Sample cup: 500 µL |
| Sample supply |
Piercing sampling |
| Compatible rack type |
ARKRAY racks (for 5 samples) |
| Operating environment |
Temperature: 10 – 30°C;
Humidity: 20 – 80% RH (No condensation) |
| Display |
20 digits × 2 lines LCD |
| Printer |
For use with 58-mm width thermal printer paper |
| Number of measurement samples |
Maximum 10 samples |
| Power requirements |
AC 100 – 240 V±10%, 50/60 Hz |
| Dimensions |
330 (W) × 515 (D) × 485 (H) mm
(Not including protrusions, eluent packs and hemolysis washing solution bottle) |