urit810测什么军用标准 (MIL-STD 810) - 介绍

Test Category Test Method & Procedure MIL-STD-810H Test Parameters High Temperature Test Method 501.7-Procedure I (A1) Duration: 7 day exposure (7 X 24 hr. cycles) Temperature: 33°C~71°C The unit is non-operational during the test. Method 501.7-Procedure II (A1) Duration: 3 day exposure (3 X 24 hr. cycles) Temperature: 32~49°C cycling temperature exposure Table 501.7-III-Procedure. High temperature cycles, climate category A1 Hot Dry The unit is operational during the test. Low Temperature Test Method 502.7- Procedure I (C1) Duration: 7 day exposure (7 X 24 hr. cycles) Temperature: -25~ -33°C The unit is non-operational during the test. Method 502.7- Procedure II (C1) Duration: 3 day exposure (3 X 24 hr. cycles) Temperature: -21~ – 32°C Low temperature cycles,Table IX. Basic climatic_C1 The unit is operational during the test. Humidity Aggravated Cycle Test Method 507.6- Procedure II Duration:10 Days Temperature: 30°C and 60°C Humidity: 95% RH, constant The unit is operational during the test. Vibration Test

Method 514.8- Procedure I

(Table514.8C-I)

Frequency 5-500Hz, Vertical rms = 1.08 g Transverse rms = 0.21g, Longitudinal rms = 0.76g Test Time: 60 minutes per axis Method 514.8- Procedure I            
(Table514.8C-VII)  Frequency 5-500Hz, Vertical rms = 2.24 g Transverse rms = 1.45g, Longitudinal rms = 1.32g Test Time: 40 minutes per axis Shock Test Method 516.8- Procedure VI Bench Handling (Drop Height  : 100 mm) The unit is operational during the test. Altitude Test Method 500.6-Procedure II Test Pressure: 15,000ft Temperature: 5°C and 40°C, both for 12 hours The unit is operational during the test.
赞(0)
未经允许不得转载:上海聚慕医疗器械有限公司 » urit810测什么军用标准 (MIL-STD 810) - 介绍